Electromagnetic flowmeter installation instructions

1. The electromagnetic flowmeter can be installed at any position on the running pipeline, and the vertical installation is preferred. When installed horizontally or obliquely, the axes of the two electrodes must be in a horizontal position
2. If the direction of liquid flow is consistent with the arrow on the nameplate of the electromagnetic flowmeter. The opposite direction of flow causes the opposite polarity.
3. It is required that the measuring tube is completely filled with liquid, and no non-full tube phenomenon is allowed.
4. There should be no ferromagnetic material close to the meter. The installation position of the electromagnetic flowmeter should be as far away as possible from the strong electromagnetic field.
5. The electromagnetic flowmeter cannot have spoilers within the upstream 5D distance. The baffle, valve or slide valve should be installed at least 2DN from the downstream side of the sensor.
6. Flanged valves cannot be connected directly to the front or back of the sensor. Because the valve also causes fluid disturbances and increases measurement errors, direct installation is not allowed under any circumstances.
7. When installing, keep the seal, grounding ring and sensor's measuring tube in a concentric position to avoid vortex flow.
8. When transporting the hoisting instrument, avoid using a tube or a sleeve to fit into the measuring tube or use a rope to hang through the measuring tube to avoid damage to the lining. Instead, the rope should be hoisted at the neck of the measuring tube.
9. In the case of serious contamination of the medium, the electromagnetic flowmeter is installed on the bypass pipe. This type of installation allows the valve to be opened and mechanically cleaned without interrupting operation.